Finite element analysis of a novel anatomical locking plate for scapular neck fracture

作者全名:"Xiang, Feifan; Xiao, Yukun; Wei, Daiqing; Tan, Xiaoqi; He, Siyuan; Luo, Liang; Yang, Yunkang"

作者地址:"[Xiang, Feifan; Xiao, Yukun; Wei, Daiqing; Yang, Yunkang] Southwest Med Univ, Dept Orthoped, Affiliated Hosp, Luzhou 646000, Peoples R China; [Xiang, Feifan; Xiao, Yukun; Wei, Daiqing; Yang, Yunkang] Sichuan Prov Lab Orthopaed Engn, Luzhou 646000, Peoples R China; [Xiang, Feifan] Macau Univ Sci & Technol, State Key Lab Qual Res Chinese Med, Macau 999078, Peoples R China; [Tan, Xiaoqi] Southwest Med Univ, Dept Dermatol, Affiliated Hosp, Luzhou 646000, Peoples R China; [He, Siyuan] Chongqing Med Univ, Dept Ophthalmol, Affiliated Hosp 2, Chongqing 400010, Peoples R China; [Luo, Liang] Chongqing Liang Jiang New Area, Dept Joint Surg, Peoples Hosp, Chongqing 401121, Peoples R China"

通信作者:"Yang, YK (通讯作者),Southwest Med Univ, Dept Orthoped, Affiliated Hosp, Luzhou 646000, Peoples R China.; Yang, YK (通讯作者),Sichuan Prov Lab Orthopaed Engn, Luzhou 646000, Peoples R China.; Luo, L (通讯作者),Chongqing Liang Jiang New Area, Dept Joint Surg, Peoples Hosp, Chongqing 401121, Peoples R China."

来源:JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH

ESI学科分类:CLINICAL MEDICINE

WOS号:WOS:000960777200001

JCR分区:Q1

影响因子:2.8

年份:2023

卷号:18

期号:1

开始页: 

结束页: 

文献类型:Article

关键词:Scapular neck fractures; Anatomical locking plate; Reconstruction plate; Biomechanical characteristics; Finite element analysis

摘要:"ObjectivesReconstruction plates (RPs) are commonly used in scapular neck fractures (SNFs): however, RPs have many defects. In this study, we evaluated a newly designed scapular neck anatomical locking compression plate (SNALCP).MethodsAn SNF finite element model (Miller-type IIB) was constructed. Plates were subsequently implanted into the scapula and fixed with screws that were grouped according to the plate used: SNALCP (A) and RP (B). Finally, loads were applied to record and analyze performance.ResultsUnder lateral, anteroposterior, and vertical compression loads, the maximum von Mises stresses on the scapula and implants of group A were smaller than those of group B. There were some differences in stress distribution between the two groups.ConclusionsSNALCP can effectively reduce the stress of the scapula and implant, making stress distribution more uniform and continuous, and has mechanical conduction advantages. Compared to RP, it provides improved stability and more reliable fixation."